V332 Pre-cleaning and maximum rinse pressures
Written V0332 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)
11.2 Rinsing/cleaning
11.2.1
When pre-cleaning is done, it is part of the reprocessing procedures.
All applicable requirements for design and maintenance of equipment included in this document should be adhered to for pre-cleaning of equipment. The maximum pressures for the dialyzer, or other limits set by the manufacturer, should be adhered to.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
ANSI/AAMI RD47:2002/A1:2003
11.2.1 Rinsing/cleaning
Many facilities pre-clean dialyzers. This process is typically accomplished with an apparatus developed by users and is intended to remove gross deposits of blood and products before rinsing and cleaning with a reprocessing machine or device.
AAMI Rationale for the Development and Provisions of this Recommended Practice
A.11.2.1 Rinsing/cleaning
Aqueous liquids rather than gases such as air are preferred for rinsing and cleaning.
Additional Guidance:
Maximum pressures allowed during reprocessing will be defined in the dialyzer manufacturer’s directions for use (DFU). Use of higher pressures may cause breaks in the dialyzer fibers and potential blood leaks. A pressure gauge must be in place to monitor the pressure of the treated water source used for rinsing the used dialyzers, and the maximum limits of the pressure to be used must be defined and known to the operator.
Use V315 if there is no pressure gauge in the water line at the rinse sink; use this tag (V332) if the manufacturer’s specified pressures are exceeded.
https://middlemolecule.com/rulebook/tags/v332/