Middle Molecule Find

Rulebook › V-tags › V339

V339 Germicidal process and rinse-out

Written V0339 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 160 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

11.4.1 Interior (blood/dialysate compartment)

11.4.1.1 Germicidal process

Chemical germicides or other procedures used for disinfecting of hemodialyzers have been shown to accomplish at least high-level disinfection when tested in dialyzers artificially contaminated with appropriate microorganisms.

If the germicide has an expiration date from the manufacturer, staff members should be sure that the chemical is not outdated. Some germicides have recommendations for maximum storage time after dilution or activation and before usage. If this is the case, the expiration date of the prepared germicide solution should be marked on the outside of the germicide solution container, and that date should be checked at the beginning of each day, before reprocessing begins. The disinfection process shall not adversely affect the integrity of the dialyzer. Germicides shall be rinsed from the dialyzer to below known toxic levels within a rinse-out period established for the particular germicide (see AAMI 12.4). To prevent injury, staff members shall take care not to mix reactive materials such as sodium hypochlorite and formaldehyde.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

ANSI/AAMI RD47:2002/A1:2003

11.4.1 Interior (blood/dialysate compartment)

11.4.1.1 Germicidal process

If formaldehyde is used as the sole germicidal agent, the CDC recommends that a concentration of 4% (W/V) be used in both the blood and dialysate compartments with a minimum contact time of 24 hours at a temperature of at least 20° C; lower concentrations or shorter contact times are appropriate if equivalent results can be demonstrated under other conditions. Formaldehyde used for reprocessing dialyzers shall not be cloudy. Concentrated formaldehyde stored under adverse conditions can polymerize to form paraformaldehyde, a white precipitate. Formaldehyde should be of United States Pharmacopoeia (USP) standards or better quality. When other germicides are used, the manufacturer’s instructions should be followed. If maximum storage temperature limitations exist, records should be maintained to document this criterion.

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.11.4.1.1 Germicidal process

Unfortunately, no realistic procedure exists whereby a dialysis center can monitor the effectiveness of the disinfection procedure. Such sophisticated microbiological tests cannot be performed in dialysis centers because they require the use of specialized equipment and highly trained microbiologists. Due to these constraints, a center should adhere strictly to established protocols for QC and QA. Tests for total bacteria and endotoxin in the water used to make up the germicide should be conducted at least monthly. If there are problems in maintaining water quality at the level established by ANSI/AAMI RD62:2001, Water treatment equipment for hemodialysis applications (referenced in RD52:2004), the testing may need to be performed more frequently.

Testing the germicide’s final-use concentration should be a part of the center’s QC program as well as verifying that each dialyzer was filled with germicide.

Potency testing of each batch of germicide is specifically recommended for batches of manually-prepared germicides, regardless of whether they are used with a manual or an automated system. Germicide solutions diluted online by automated machines should be checked for concentration at least monthly.

CMS requires (at 42 CFR § 494.50) that dialyzers not be subjected to multiple germicide solutions because of possible combined actions of the germicides on the hemodialyzer membrane. This requirement does not apply to the original sterilization process or chemical cleaning agents to which the hemodialyzer may be exposed for short periods during the cleaning process for reuse.

Additional Guidance:

The reuse staff must be knowledgeable about the germicide used and the risks this germicide presents to him/her and to the patients, such as injuries that may be caused by mixing reactive materials such as sodium hypochlorite and formaldehyde.

Containers of germicide should be dated to indicate dilution and discard dates.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 160 (the full 469-page PDF) before quoting it.