V340 Filling with germicide: at least 90%
Written V0340 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)
11.4.1.4 Chemical germicidal procedure
If applicable, the hemodialyzer shall be filled with the germicide solution until the concentration in the hemodialyzer is at least 90% of the prescribed concentration.
The ports of chemically disinfected dialyzers shall be disinfected and then capped with new or disinfected caps. The caps may be disinfected with dilute bleach, with the chemical used for disinfecting the hemodialyzer, or with any other germicide approved by the FDA as a disinfectant that does not adversely affect the materials of the dialyzer.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
If a manual reprocessing system is in use, the blood and dialysate compartments must be filled with a volume of germicide equal to three times the total volume of the blood and dialysate compartments of the dialyzer (to equal three compartment volumes) in order to reach at least 90% of the prescribed germicide concentration. Reuse logs must include documentation of verification of the desired (prescribed) concentration of germicide.
Used and new port caps must be disinfected prior to use. The reuse technician should be knowledgeable about the minimum contact time required for germicide disinfection of port caps.