V342 Cleaning the dialyzer exterior

Written V0342 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 163 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)

11.4.2 Exterior: low-level disinfection

The outside of the dialyzer should be soaked or wiped clean of visible blood and other foreign material. For chemically disinfected dialyzers, a low-level germicide that is compatible with the dialyzer’s materials of construction should be used for this purpose.

CMS's interpretive guidance

Interpretive Guidance 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.11.4.2 Exterior

Sodium hypochlorite at a concentration of 0.05% is usually suitable [for external cleaning]. Certain commercial low-level disinfectants may cause some plastics used for dialyzers to crack after repeated or prolonged exposure.

Low-level germicides satisfactorily clean the exterior of the device to a degree comparable with what a new dialyzer receives. For example, 1:100 dilution of household bleach will achieve the concentration of sodium hypochlorite specified [here].

Additional Guidance:

The exterior of each dialyzer must be cleaned after reprocessing steps are complete. Spraying the dialyzer with germicide is generally unsatisfactory, unless all the surfaces of the dialyzer are covered with the spray. Dialyzers may be dipped or allowed to soak in a germicide solution, or wiped with a disposable cloth saturated with a germicide solution.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 163 (the full 469-page PDF) before quoting it.