V341 Checking germicide concentration

Written V0341 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 162 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

11.4.1.6 Chemical germicide concentration

[For] reprocessing systems in which each batch of germicide is manually prepared, each batch of germicide shall be tested before use to verify the proper concentration of the germicide. This requirement does not apply in cases in which each dialyzer is tested for concentration before setup.

When the germicide is diluted on-line, its concentration in the hemodialyzer immediately after reprocessing should be checked at least monthly for each reprocessing system.

When the germicide is partially or fully diluted by the user, the solution [shall] be thoroughly mixed [before use].

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance 494.50(b)(1)

The system in use must include verification of germicide concentration in the dialyzers after reprocessing.

If the germicide is diluted by the user (batch), the germicide manufacturer's instructions for dilution must be followed, and the solution must be thoroughly mixed before use. If each dialyzer is tested for germicide concentration prior to rinsing, it is not necessary to test the germicide concentration before use of the batch.

If the germicide is diluted on-line, the concentration in a dialyzer from each reprocessing system must be checked immediately after reprocessing at least monthly.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 162 (the full 469-page PDF) before quoting it.