V345 Storage of reprocessed dialyzers

Written V0345 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 166 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)

11.7 Storage

Reprocessed dialyzers that meet the performance and inspection criteria for multiple use should be stored according to the provisions of [AAMI] 8.2. Prolonged storage (greater than 1 month) should be documented to be safe and effective.

Dialyzers that have exceeded the facility’s maximum storage time shall be reprocessed or discarded. The dialyzer and disinfectant labeling should be consulted regarding proper storage conditions.

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

Additional Guidance:

The AAMI RDD Committee were, unaware of any adverse effects of storage for up to 1 month. Dialysis facilities that store its multiple use dialyzers for more than one month should have documented evidence that such a time period does not pose a risk to patient safety..

There might be occasions when reprocessed dialyzers are stored for extended periods of time, such as when patients are absent (e.g. hospitalized, vacation). There must be a system for ensuring dialyzers are not stored longer (without reprocessing) than the maximum time limit specified by the germicide manufacturer and facility policy.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 166 (the full 469-page PDF) before quoting it.