V346 Written procedure for germicide residue tests
Written V0346 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)
12 Preparation for dialysis and testing for chemical germicides and potentially toxic residues
A written procedure that has been shown to be effective shall be followed.
12.5 Written procedure for tests for germicide or other residues There shall be a written procedure for all tests employed in preparing the dialyzer for use, including mention of each test’s sensitivity. The germicide manufacturer’s instructions for use should be consulted in determining the maximum residual level. The physician in charge of the reuse program shall approve any alterations in the procedures.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
AAMI Rationale for the Development and Provisions of this Recommended Practice
A.12 Preparation for dialysis and testing for chemical germicides and potentially toxic residues
When the [AAMI RDD] Committee revised RD47 in 2002, it decided that sufficient information was available to indicate that the residual level of formaldehyde should be reduced to less than 3 ppm. The testing technology for residual formaldehyde had also improved, and it was feasible to easily test to less than 3 ppm.
The [AAMI RDD] Committee considered establishing maximum residual levels for germicides other than formaldehyde. Because these newer germicides are all cleared by the FDA and could have different allowable levels of residuals even for the same generic type of germicide, [they] determined that it is best to recommend that the manufacturer’s instructions for use be followed. The [AAMI RDD] Committee noted that toxicology studies are favorable for some of these agents, and the FDA reviews labeling information for them, which includes the maximum residual level.
When checking for the presence or concentration of the germicide in the hemodialyzer, [do not place anything] into the blood or dialysate ports of the device (e.g., test strip or syringe) to withdraw the sample. Doing so may damage the fibers of the dialyzer and lead to blood leaks during dialysis. . , If a germicide test strip or kit is being used, the instructions provided by the manufacturer should be followed.
Additional Guidance:
The reagents used for the germicide tests must be sensitive to the levels specified by the germicide manufacturer (i.e. high level for presence test, low level for residual test).