V350 Presence test of each dialyzer, if the maker requires it
Written V0350 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)
12.3.1 Presence test of each hemodialyzer
The presence of germicide may be ensured by directly testing each dialyzer for the presence of germicide. Certain germicide manufacturers require this testing before the germicide is rinsed from the hemodialyzer, and these instructions should be followed.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
Manufacturers of peracetic acid and glutaraldehyde require every dialyzer be tested for presence of a sufficient concentration of germicide after storage and before use.
The test for germicide presence (at the potency recommended by the germicide manufacturer) should be done before rinsing and priming.