V349 Germicide contact time and storage record
Written V0349 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)
12.3 Verification of germicidal contact
The contact time of the germicide or disinfection procedure shall comply with the facility’s protocol and the manufacturer’s recommendations.
The presence of chemical germicide in each hemodialyzer shall be ensured through either direct testing or an on-line process and procedural control. If other disinfection procedures (e.g., heat) are used, there shall be methods to ensure that each hemodialyzer has been properly subjected to the disinfection process. A record shall be kept indicating that the dialyzer has undergone the appropriate storage time, and the record shall be appropriately verified.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
Staff must be able to determine the date and time of each dialyzer’s last reprocessing. Responsible staff (e.g., reuse technician, direct care staff) must ensure the dialyzers have been exposed to the germicide for the required contact time before set up and use.
For systems using heat disinfection, staff must verify the dialyzer was exposed to the appropriate temperature for the required time period to ensure disinfection.