Middle Molecule Find

Rulebook › V-tags › V352

V352 Priming and rinsing out germicide

Written V0352 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 171 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)

12.4 Priming the dialyzer and rinsing the germicide

If the manufacturer’s instructions so require, a germicide presence test shall be performed before the germicide is rinsed from the dialyzer.

The dialyzer shall be rinsed and primed according to a written procedure that has been documented to produce a reduction in the concentration of germicide to an acceptable level and result in a physiological solution in the blood and dialysate compartments. The dialyzer manufacturer’s instructions should be considered in developing these procedures.

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

The dialyzer rinsing/priming and preparation for use procedures followed must be in accordance with the dialyzer manufacturer’s requirements for the germicide in use.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 171 (the full 469-page PDF) before quoting it.