V352 Priming and rinsing out germicide
Written V0352 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)
12.4 Priming the dialyzer and rinsing the germicide
If the manufacturer’s instructions so require, a germicide presence test shall be performed before the germicide is rinsed from the dialyzer.
The dialyzer shall be rinsed and primed according to a written procedure that has been documented to produce a reduction in the concentration of germicide to an acceptable level and result in a physiological solution in the blood and dialysate compartments. The dialyzer manufacturer’s instructions should be considered in developing these procedures.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
The dialyzer rinsing/priming and preparation for use procedures followed must be in accordance with the dialyzer manufacturer’s requirements for the germicide in use.