V353 Residual germicide test before dialysis

Written V0353 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 172 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)

12.4.1 Testing for residual germicide

Residual germicide shall be measured by a test of appropriate sensitivity according to a written procedure to ensure that the germicide level is below the maximum recommended residual concentration . . . Completion of this step shall be documented, along with the signature or other unique means of identifying the person performing the test.

A written policy should establish the maximum allowable time between rinsing the germicide from the dialyzer and beginning dialysis . . . The priming, removal, and residual testing process should be reinstituted after [any] delay sufficient to bring concentrations of germicide above the recommended level (rebound) [by elution from these compartments]. Additional rinsing should be performed to yield a germicide level below the maximum recommended concentration before initiating of dialysis.

A rinse procedure should be defined and documented step by step, and all personnel should be familiar with and follow it.

If heat disinfection is used, the dialyzer should be cool to the touch before it is primed with saline.

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.12.4.1 Testing for residual germicide

Because certain germicides have been demonstrated to disperse into the solid components or less rapidly exchangeable compartments of the of the hemodialyzer, the priming, removal, and residual testing process should be reinstituted after any delay that provides sufficient time for concentrations of germicide to rebound above the recommended level by elution from these compartments.

A number of procedural steps have been identified that, if not followed, may cause residual germicide to remain in the hemodialyzer following rinsing. The following list of instructions, though not all inclusive, should be carefully considered when developing a facility’s rinsing procedure:

  • Air bubbles in the fibers can cause individual fibers to become blocked. Be sure that the arterial line is fully primed before [it is connected] to the hemodialyzer. If using peracetic acid-type germicide, [the blood side should be flushed] before beginning dialysate flow.
  • Air trapped in the dialysate side of the hemodialyzer may cause germicide to also remain trapped in portions of the hemodialyzer. [The dialyzer should be rotated] during the rinsing process. This action normally releases the trapped air and allows the germicide to be fully rinsed.
  • Germicide may back up into the heparin or monitor lines. [The heparin line must be] clamped [so] that fluid is not forced into the monitor lines.
  • Germicide may back up into the saline bag during the rinsing procedure. [Procedures for initiation of treatment must account] for all situations that may force fluid from the dialysis circuit back into the saline bag.
  • Take care to avoid a false-negative residual disinfection test, which can happen if you sample too quickly after a quantity of saline has been infused.
  • Discard the prime solution when beginning blood flow to the hemodialyzer.

Additional Guidance:

If the extracorporeal circuit is prepared ahead of time, staff should repeat the residual germicide test just prior to treatment initiation to allow detection of any “rebound” of germicide. This is particularly a concern if fluid circulation through the circuit and dialyzer is stopped, as may happen when a patient is late coming to treatment.

If facility practice is to discard the priming solution by “bleeding patients on,” policy and practice should reflect a requirement that a staff member constantly attend that patient while the venous line is open as blood fills the extracorporeal circuit, to prevent accidental blood loss.

Generally, manufacturers of dialyzers labeled for reuse address the need to discard the prime solution by advising users to replace the saline bag used for priming and refresh all fluid in the extracorporeal circuit with saline from the new bag prior to beginning patient treatment.