Middle Molecule Find

Rulebook › V-tags › V417

V417 Life Safety Code, no grade-level exit

Written V0417 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.60(d)(1) · Appendix H Rev. 246, page 192 · checked 12 Nov 2026 · guidance updated Aug 17, 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§ 494.60(d) - Standard: Fire safety. (1) Except as provided in paragraph (d)(2) of this section, dialysis facilities that do not provide one or more exits to the outside at grade level from the patient treatment area level, must comply with provisions of the 2012 edition of the Life Safety Code of the National Fire Protection Association (NFPA 101 and tentative interim amendments TIA 12-1, TIA 12-2, TIA 12-3, and TIA 12-4) applicable to Ambulatory Health Care Occupancies [which is incorporated by reference in paragraph (f) of this section], regardless of the number of patients served.

Appendix H quotes this text a little differently from the regulation as published today. The eCFR has the current wording.

Read 42 CFR 494.60 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.60(d)(1)

ESRD facilities that do not provide one or more exits to the outside at grade level from the patient treatment area must comply with the requirements contained in the 2012 edition of the Life Safety Code (NFPA 101) and the Tentative Interim Amendments TIA 12-1, TIA 12-2, TIA 12-3 and TIA 12-4 applicable to the Ambulatory Health Care Occupancies, regardless of the number of patients served.

Dialysis facilities that provide one or more exits to the outside at grade level would be exempt from meeting the requirements in the 2012 edition of the Life Safety Code (NFPA 101) and the applicable tentative interim amendments. Exempt facilities must continue to comply with state and local fire codes.

“Exit to the outside at grade level from the patient treatment area level” is to mean that ESRD facilities are on the ground or grade level of a building where patients do not have to traverse up or down stairways to evacuate to the outside. Accessibility ramps in the exit area that provide an ease of access between the patient treatment level and the outside ground level are not considered stairways.

A dialysis facility required to comply with the LSC must operate in accordance with the code chapters (Chapters 20 or 21) applicable to ambulatory health care occupancies. However, a dialysis facility may be located in a mixed occupancy building. If a dialysis facility is located in a building with other tenants and occupancies, it must be separated by a one-hour fire-resistance-rated wall in accordance with the LSC. For purposes of this regulation, when a portion of the dialysis facility is used intermittently by another entity (e.g., an exam room is used for the nephrologist’s office practice) it is not intended that the portion used intermittently would be separated from the dialysis facility by a one-hour fire wall.

If the dialysis facility is located within a hospital, but not separated from the hospital by two-hour fire wall construction, the dialysis facility must meet the hospital LSC requirements. If a hospital-based chronic outpatient facility provides acute services for hospitalized patients in the same space and within the hospital walls, the outpatient dialysis facility must meet the more-stringent hospital chapters of the LSC.

Survey instructions found in State Operations Manual (SOM), Appendix I, are used as applicable, along with the Fire Safety Survey Report Form, 2786U to survey for fire safety. The fire safety standard will be surveyed by a fire specialist from the State Agency or a CMS-approved Accrediting Organization, if accredited; generally, the LSC inspections will be done separately from the rest of the ESRD facility survey.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 192 (the full 469-page PDF) before quoting it.

Connected tags

Tags whose guidance points here