V407 HD patients in view of staff

Written V0407 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.60(c)(4) · Appendix H Rev. 246, page 192 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§ 494.60(c)(4) Patients must be in view of staff during hemodialysis treatment to ensure patient safety, (video surveillance will not meet this requirement).

Read 42 CFR 494.60 on the eCFR

CMS's interpretive guidance

Interpretive Guidance §494.60(c)(4)

Each patient, including their face, vascular access site, and bloodline connections, must be visible to a staff member throughout the dialysis treatment. Allowing patients to cover access sites and line connections provides an opportunity for accidental needle dislodgement or a line disconnection to go undetected. This dislodgement or disconnection could result in exsanguination and death in minutes.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 192 (the full 469-page PDF) before quoting it.