V180 Conventional dialysate: bacteria and endotoxin

Written V0180 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 44 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

4.3.2.1 Bacteriology of conventional dialysate

Conventional dialysate should contain a total viable microbial count lower than 200 CFU/mL and an endotoxin concentration of lower than 2 EU/mL.

The action level for the total viable microbial count in conventional dialysate should be 50 CFU/mL and the action level for the endotoxin concentration should be 1 EU/mL. If levels exceeding the action levels are observed in the dialysate, corrective measures, such as disinfection and retesting, should promptly be taken to reduce the levels.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

AAMI Rationale for the Development and Provision of This Recommended Practice

A.4.3.2.1 Bacteriology of conventional dialysate

It is now clear that endotoxins, endotoxin fragments, or other bacterial products cross at least some dialysis membranes under some operating conditions.

In addition to the risk of acute pyrogenic reactions, indirect evidence increasingly shows that chronic exposure to low amounts of endotoxin may play a role in some of the long-term complications of hemodialysis therapy. Patients treated with ultrafiltered dialysate have demonstrated a decrease in serum β2-microglobulin concentrations, a decrease in markers of inflammation, and an increased responsiveness to erythropoietin. In longer-term studies, use of microbiologically ultrapure dialysate has been associated with a decreased incidence of β2-microglobulin-associated amyloidosis, better preservation of residual renal function, and improved nutritional status. For those reasons, the AAMI RDD Committee reduced the recommended maximum microbial count in the dialysate to 200 CFU/mL and added a recommendation that the endotoxin concentration not exceed 2 EU/mL. The values are the same as those for water used to prepare the dialysate (ANSI/AAMI RD62:2001), implying that the dialysate proportioning system should not add significantly to the microbiological burden in the water. Although the AAMI RDD Committee did not review supporting data, it considered contemporary dialysate delivery systems to be fully capable of performing at this level, provided that the user followed the manufacturer’s instructions on cleaning and disinfecting the system, including disinfection of the line between the water distribution system and the concentrate mixing chambers of the dialysate proportioning system.

Additional Guidance:

Conventional dialysate refers to the dialysate typically used for hemodialysis in the U.S., as opposed to “ultrapure dialysate.” Recognize that the purity of dialysate is essential, in that “reverse” ultrafiltration can occur, allowing dialysate to cross the dialyzer membrane and enter the patient’s bloodstream. This can occur with most dialyzers, especially high-flux dialyzers at the distal end of the dialyzer, and especially if the patient does not have much fluid weight to remove (allowing less ultrafiltration pressure to be used). A minimum ultrafiltration rate (UFR), as per the dialysis machine manufacturer’s DFU, should be maintained to prevent reverse ultrafiltration.

It is not expected that concentrates would be cultured or tested for endotoxin levels. Machine cultures (of dialysate) are used as evidence to determine if this requirement is met.

“Promptly” may be met if action is taken within 48 hours of receiving the results of testing. One example of a corrective measure is to repeat cultures, particularly in cases where one of several cultures exceeds the action level, or to disinfect the system and repeat cultures. Action should also include notifying the medical director of the results.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 44 (the full 469-page PDF) before quoting it.