V179 Water bacteriology: at installation and after
Written V0179 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
4.1.2 Bacteriology of water
The manufacturer or supplier of a complete water treatment and distribution system [must] demonstrate that the complete water treatment, storage, and distribution system is capable of meeting these requirements at the time of installation.
Following the installation of a water treatment, storage, and distribution system, the user is responsible for continued monitoring of the water bacteriology within the system and for complying with the requirements of this standard, including those related to action levels.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
While this standard is primarily addressed to the manufacturer of equipment, each dialysis facility’s medical director is responsible for the quality of medical care provided to patients. The medical director oversees the installation and monitoring of the water treatment and distribution system. Their failure to evaluate and validate the environment where devices and equipment will be used may result in non-compliance with this requirement.
Existing facilities must monitor AAMI water chemical analysis and microbial testing and take action if the results fall outside AAMI standards. If culture results exceed the action levels, the facility may need to repeat cultures or disinfect the system and repeat cultures (depending on the number of positive cultures, etc.) and continue treatment while awaiting the results.
AAMI recommendations at 4.1.1, Table 1, recommend maximum levels of contaminants above which corrective measures should be taken. Testing should occur initially upon installation, and at annual intervals thereafter. If the water supply for the facility is from a private well, an annual analysis of the product water quality may not be sufficient to ensure that the feed water requirements of the water treatment system in use are continuously met. The quality of water from the well may change over time, and private wells are not routinely monitored. More frequent analysis may be needed if the well is subject to seasonal changes or contamination from sources such as septic tanks, underground fuel storage tanks, or agricultural waste and chemicals. When more frequent than annual testing is performed by the dialysis facility, testing may be limited to specific contaminants that are known to be of concern.
Frequency and methods for monitoring water bacteriology are addressed at V252 & V255.