V230 Cleaning and disinfecting mixing equipment
Written V0230 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
6.4.1 Mixing systems (cleaning)
Concentrate mixing equipment should be either: (1) completely emptied, cleaned, and disinfected according to the manufacturer’s instructions; or (2) cleaned and disinfected using a procedure demonstrated by the facility to be effective in routinely producing concentrate [meeting the standards related to allowable bacterial and endotoxin levels].
The mixing and disinfection data should be recorded for each mix and disinfection cycle using a dedicated log.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
The disinfection log may be kept electronically or on paper and may be for only this purpose or inclusive of other operations. Mixing and disinfection data may be maintained separately or jointly. Guidance for recordkeeping of mixing systems and maintenance of the mixing log is found at V229.
Standards related to allowable bacterial and endotoxin levels, i.e. bacteriology of conventional dialysate, is found at V180.
https://middlemolecule.com/rulebook/tags/v230/