V187 Schematic diagrams and labeled piping and components

Written V0187 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 50 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

8Environment (schematic diagrams and labels)

Water systems should include schematic diagrams that identify components, valves, sample ports, and flow direction.

Additionally, piping should be labeled to indicate the contents of the pipe and direction of flow.

If water system manufacturers have not done so, users should label major water system components in a manner that not only identifies a device but also describes its function, how performance is verified, and what actions to take in the event performance is not within an acceptable range.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

The use of text labels, such as “RO Water,” and color-coded “arrow tape” provide a convenient means of identifying pipe content and flow direction.

An example of labeling for a regenerable softener is given below.

Additional Guidance:

There must be a schematic diagram which allows the staff to follow the flow of the water through the components and each component and the piping must be labeled as described in the recommended practice for the Environment where the water purification and storage system is located.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 50 (the full 469-page PDF) before quoting it.