V188 Sediment filters: gauges, pressure drop, backwash timer
Written V0188 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.2.2 Sediment filters (equipment configuration)
6.2.2 Sediment filters (monitoring)
5.2.2 Sediment filters:
Bed filters should be fitted with gauges to measure the hydrostatic pressure at the filters’ inlet and outlet.
6.2.2 Sediment filters:
Sediment filters should be monitored on a periodic basis… [for a] pressure drop (∆P) across the filter [that] can be used to determine when the filter is retaining particulate matter to the point that the filter will no longer allow the required water flow without an excessive reduction in pressure at the outlet of the filter. A backwash cycle is used to remove particulate matter from the sediment filter. The frequency of backwashing should follow the manufacturer’s recommendations.
Sediment filter monitoring should include daily verification that the timer used to initiate backwashing cycles is set to the correct time of day. A log sheet should be developed to record the pressure drop measurements and timer verifications.
[Refer to RD62:2001, 4.3.8 Sediment filters:] Sediment filters shall have an opaque housing or other means to inhibit proliferation of algae.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.2.2 Sediment filters
Permanent, back-washable sediment filters, also known as “bed filters,” are frequently located at or near the beginning of hemodialysis water treatment systems and are intended to remove relatively coarse particulate materials from incoming water. Although a single filtration medium may be used, bed filters known as multimedia filters are more commonly selected. These units contain multiple layers, each layer retaining progressively smaller particles. In this way, the bed is used to its fullest extent; the largest particles are removed in the first layer contacted by the water, and the smallest in the final layer.
As the bed accumulates particulate material, open passages begin to clog, and resistance to the water flowing through the filter increases. Ultimately, the increased resistance to flow will result in a reduction of water supply to downstream components. To prevent this situation from occurring, bed filters are cleaned by periodic backwashing, which is accomplished either manually or by using a timer-activated control valve.
Bed filters should be fitted with gauges to measure the hydrostatic pressure at the filters’ inlet and outlet. These values can be used to determine the dynamic pressure drop across the filter (delta pressure or ΔP), which serves as an index of resistance to flow and provides a basis for setting the frequency of backwashing.
Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4 include:
- For sediment filters, monitor the pressure drop across the filter daily, looking for a pressure drop of less than a number determined by facility policy.
- For the sediment filter backwashing cycle, monitor the backwash cycle timer setting daily at the beginning of the day, looking for the backwash clock set to the time determined by facility policy.
Additional Guidance:
Sediment filters are not required in every facility; instead, the source water should determine the necessary water treatment components. If sediment filters are in use, the facility must follow these requirements.
If a water treatment system includes multiple components that backwash, the “time” set on each timer may need to be staggered to allow sufficient water to be available for the backwashing. This may result in some timers being set an hour or two different from the correct time. If so, there should be a posted notice to that effect.
Pressure readings should be taken while the equipment is running. To determine the pressure drop (or delta P), observe the pressure readings on the gauges before and after the filter. If the gauge before the filter reads 70 mmHg and the gauge after the filter reads 50 mmHg, the pressure drop or “Delta P (ΔP)” is 20. The ΔP must be within the facility set limits; a result higher than the limits indicates that the filter needs to be backwashed or replaced.