V189 Cartridge filters: pressure drop and replacement
Written V0189 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40 (a)
5.2.3 Cartridge filters (equipment configuration)
6.2.3 Cartridge filters (monitoring)
5.2.3 Cartridge filters
The cartridge is contained within an opaque filter housing with seals to separate the feed and product water streams.
When the maximum [pressure drop] ∆P recommended by the filter manufacturer is reached, the cartridge should be replaced according to the manufacturer’s instructions.
6.2.3 Cartridge filters
Cartridge filters should be monitored on a periodic…. basis for a [pressure drop] ∆P across the filter [that] can be used to determine when the filter is retaining particulate matter to the point that the filter will no longer allow the required water flow without an excessive reduction in pressure at the outlet of the filter. A marked decrease in ∆P without a corresponding decrease in flow rate may indicate a loss of filter integrity. Follow the manufacturer’s recommendations concerning when to replace cartridge filters.
Replacement of the cartridge will usually be indicated by an increase in ∆P to some specified value. A log sheet should be developed to record the pressure drop measurements.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.2.3 Cartridge filters
Cartridge filters consist of a cylindrical cartridge of the filter medium with a central drainage core. Although cartridge filters may be installed at the inlet to a water system, their usual application is as a final filtration step before reverse osmosis.
As the cartridge accumulates particulate material, resistance to flow through the filter increases, as indicated by an increase in ∆P.
Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4 include: For cartridge filters: monitor for pressure drop across the filter daily, looking for a pressure drop less than a number determined by facility policy. Since some values are system-specific, the facility should define an acceptable operating range for monitoring pressure drops based on the manufacturer’s instructions or measurements of system performance.
Additional Guidance:
Cartridge filters are not required in every facility; the source water should determine the necessary water treatment components. If cartridge filters are in use, the facility must follow these requirements.
A marked decrease in ΔP could indicate that there is no filter in the container.