V194 Carbon: iodine number of at least 900, no regenerated carbon

Written V0194 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 59 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.2.5 Carbon adsorption: (Iodine #900/replacement)

When granular activated carbon is used as the media, it shall have a minimum iodine number of 900. Other forms of carbon should not be used unless there is performance data to demonstrate that each adsorption bed has the capacity to reduce the chloramine concentration in the feed water to less than 0.1 mg/L when operating at the maximum anticipated flow rate for the maximum time interval between scheduled testing of the product water for chloramines.

Regenerated carbon shall not be used for hemodialysis applications.

Refer to RD62:2001, 4.3.9 Carbon adsorption media: Exhausted carbon adsorption media shall be discarded and replaced with new media according to a replacement schedule determined by regular monitoring.

CMS's interpretive guidance

Interpretive Guidelines §494.40(a)

ANSI/AAMI RD62:2001

4.3.9 Carbon adsorption media

For example, when testing between the beds shows that the first bed is exhausted, the second bed should be moved into the first position, the second bed replaced with a new bed, and the exhausted bed discarded.

Other forms of carbon should not be used unless there is performance data to demonstrate that each adsorption bed has the capacity to reduce the chloramine concentration in the feed water to less than 0.1 mg/L when operating at the maximum anticipated flow rate for the maximum time interval between scheduled testing of the product water for chloramines.

ANSI/AAMI RD52

5.2.5 Carbon adsorption

Some granular activated carbon contains aluminum, which can elute from the carbon and add to the burden of aluminum to be removed by reverse osmosis or ion exchange. The use of acid-washed carbon minimizes this source of aluminum in the water.

Additional Guidance:

Facilities using an exchange tank system should establish a schedule for replacing the exhausted carbon adsorption media based on their experience with these tanks to prevent interruptions to patient services. The date of exchange of tanks should be documented on the tank and in the appropriate log. Facilities using back-washable carbon systems should establish a schedule for back-washing the tanks and document the system's function. Back-washing does NOT regenerate the carbon; it rearranges the carbon in the tank, exposing sites that have not yet been used to adsorb chlorine or chloramine. Back-washable systems do exhaust; the responsible staff member should be able to describe how they will replace the carbon when indicated. A “schedule” could refer to “every X months” rather than a specific date or month and should be based on past experience at the facility. Carbon in the tanks can be removed and replaced (“rebedding”) on site when the tanks are offline.

Granulated activated carbon (GAC) with a minimum iodine number of 900 must be specified when replacement carbon is ordered. Acid-washed carbon is recommended, as it will protect the RO from exposure to excess aluminum, but is not required as the RO would still protect the patient.