V195 Carbon beds: at least 5 minutes' contact time each
Written V0195 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.2.5 Carbon adsorption: (10 minute EBCT)
When granular activated carbon is used as the [adsorption] media, it shall have a minimum iodine number of 900, and each bed shall have a minimum empty bed contact time of 5 minutes at the maximum flow rate through the bed.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
The empty bed contact time (EBCT) of the granulated activated carbon (GAC) system should be periodically calculated for the maximum water flow through the carbon tanks. The facility determines the frequency of monitoring based on the manufacturer’s directions for use. Specific facility conditions may warrant more frequent monitoring, such as chlorine disinfection practices for source water. Water flow rates may vary, altering the need for more or less GAC to achieve the 10-minute total EBCT (for 2 beds). If additional patient treatments or shifts are added, the resultant increased water demand should prompt the medical director and technical staff to consider adding more carbon in order to maintain the minimum EBCT. “Each adsorption bed” refers to the primary tank or tanks as one adsorption bed and the secondary tank or tanks as another adsorption bed.
https://middlemolecule.com/rulebook/tags/v195/