V196 Chlorine/chloramine testing: timing and method

Written V0196 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 60 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

6.2.5 Carbon adsorption (monitoring and testing frequency)

Testing for free chlorine, chloramine, or total chlorine should be performed at the beginning of each treatment day prior to patients initiating treatment and again prior to the beginning of each patient shift. If there are no set patient shifts, testing should be performed approximately every 4 hours.

Results of monitoring of free chlorine, chloramine, or total chlorine should be recorded in a log sheet.

Testing for free chlorine, chloramine, or total chlorine can be accomplished using the N.N-diethyl-p-phenylene- diamine (DPD) based test kits or dip-and-read test strips. On-line monitors can be used to measure chloramine concentrations. Whichever test system is used, it must have sufficient sensitivity and specificity to resolve the maximum levels described in [AAMI] 4.1.1 (Table 1) [which is a maximum level of 0.1 mg/L].

Samples should be drawn when the system has been operating for at least 15 minutes. The analysis should be performed on-site, since chloramine levels will decrease if the sample is not assayed promptly.

CMS's interpretive guidance

Interpretive Guidelines § 494.40(a)

ANSI/AAMI RD52:2004

5.2.5 Carbon adsorption

In addition to removing free chlorine and chloramine, carbon also adsorbs a wide variety of other substances, including both naturally occurring and synthetic organic compounds. The capacity of carbon to remove free chlorine and chloramine may be reduced when other substances “mask” reactive sites on the carbon media. Additionally, the efficiency of free chlorine and chloramine removal is reduced as the pH increases or the temperature decreases. The net effect of those variables is that the finite capacity of carbon beds to remove free chlorine and chloramine cannot be predicted with any certainty. Therefore, their performance needs to be monitored frequently.

6.2.5 Carbon adsorption

Carbon adsorption performance is monitored by measuring free chlorine and/or chloramine concentrations in the water exiting the first carbon bed of a series-connected pair. It should be noted that sampling for total chlorine (the sum of free chlorine and chloramine), allowing a maximum level of 0.1 mg/L of total chlorine, is often simpler than analyzing for free chlorine and chloramine separately.

More frequent monitoring may be appropriate during temporary operation with a single carbon bed, which can occur following breakthrough of the first bed. In such instances, testing is performed on water exiting the second carbon bed in a series-connected pair. The decision to increase the frequency of monitoring should be based on the system's past performance and whether changes in feed water quality have occurred.

AAMI Rationale for the Development and Provision of This Recommended Practice

A.6 Monitoring

A.6.2.5 Carbon adsorption

Intensive monitoring of carbon adsorption beds is recommended due to the long history of adverse events associated with chloramine contamination of dialysate. Chloramine concentrations in municipal water can fluctuate from day to day, and the capacity of carbon adsorption beds to remove chloramine can vary depending on the pH and temperature of the water, the nature of the chloramine compounds present, and the presence of other substances in the water. The dependence of chloramine removal on multiple factors makes the performance of carbon adsorption beds unpredictable. Patient safety can only be ensured by intensively monitoring the performance of the carbon adsorption bed. Configuring carbon adsorption beds in series and sampling from a port located between the two beds provides one margin of protection against chloramine breakthrough. When chloramine is first detected in the effluent from the first adsorption bed, essentially the full capacity of the second bed remains available for chloramine removal. This reserve capacity allows the user to conveniently replace the exhausted bed without risk to patients. The exhausted bed is discarded, the second bed is moved into the first position, and a new bed is placed in the second position. A new bed of virgin carbon shall be used for replacement. Carbon cannot be regenerated in a dialysis facility, and the use of regenerated carbon is prohibited by ANSI/AAMI RD62:2001 (see 2.3 in that AAMI document). Backwashing of carbon beds does not regenerate the carbon, although it may allow more efficient use of the bed’s capacity by removing channels that can form in the bed during routine operation.

AAMI Rationale for the Development and Provision of This Recommended Practice

A.6.2.5 Carbon adsorption

The recommendation that the water purification system operate for at least 15 minutes before sampling is to guard against inadvertently drawing water that has been in the bed for an extended period.

Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4, include: For carbon adsorption beds, monitor the product water levels of free chlorine and/or total chlorine between the beds prior to beginning each patient shift. The expected result is less than or equal to 0.1 mg/L of total chlorine.

Additional Guidance:

For parallel-connected tanks/beds, testing must be performed for each set of tanks/beds each time testing is conducted.

Test strips with color comparison charts that indicate a low-level reading of zero and a first “number” of 0.5 are not sufficiently sensitive to detect levels as low as 0.1 and should not be used for testing product water for safe levels of chlorine/chloramine. An indication of “0” on the comparison charts does not suffice to demonstrate that the strips are sensitive to “0.” Consult the manufacturer’s guidance or contact the manufacturer if there are any questions regarding the sensitivity of specific test strips. When deciding whether to use a “quantitative” or “qualitative” test methodology, it is essential to recognize that determining low levels of chlorine (i.e., <0.1 ppm) requires the use of the quantitative method.

If an online chlorine/chloramines monitor is in use that incorporates an automated alarm, particular testing times are not required. Facility policy and practice should follow the manufacturer’s guidance regarding any required comparison testing and calibration of the monitor.

The ability to discern colors is a crucial job function for individuals responsible for interpreting colorimetric tests. Depending on the test method used, staff assigned this responsibility must be capable of properly interpreting the results and accurately distinguishing and identifying any colorimetric results; alternatively, a digital meter must be used. If a digital meter is used, it should be calibrated to zero before testing to ensure accurate measurements.

For deficient practices with exceeding the acceptable level of chlorine or chloramine refer to V270-273 of this section.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 60 (the full 469-page PDF) before quoting it.