V197 Breakthrough at the first carbon bed: up to 72 hours

Written V0197 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 63 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.2.5 Carbon adsorption: (positive sample of chlorine or chloramine)

When samples from the first sampling port are positive for chlorine or chloramine, operation may be continued for a short time (up to 72 hours) until a replacement bed is installed, provided that samples from the second sampling port remain negative. The replacement bed should be placed in the second position, and the existing second bed should be moved to the first position to replace the exhausted bed. If it is not possible to rotate the position of the beds, both beds should be replaced.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

AAMI Rationale for the Development and Provision of This Recommended Practice

A.5.2.5 Carbon adsorption

The AAMI RDD Committee recognized that it might not be practical to rotate the bed positions in installations that use large, backwashable carbon beds. However, there was concern that the capacity of the second bed might decrease unpredictably and no longer provide adequate backup if a breakthrough occurred in the first bed. For this reason, the AAMI RDD Committee recommended replacing both beds if bed rotation was not possible.

Additional Guidance:

When facilities operate with one exhausted carbon bed for up to 72 hours, the log of testing should include the actual times testing was done rather than indicating “1st, 2nd, or 3rd” shift.

A “negative” sample is interpreted as a total chlorine (the sum of free chlorine and chloramine) level less than or equal to 0.1mg/L. A “positive” sample is interpreted as a total chlorine level above 0.1mg/L.

Note: When separately analyzing the removal of free chlorine and chloramines, a maximum level for free chlorine removal is 0.5 mg/L, and removal of chloramine to a maximum level of 0.1mg/L is necessary to protect hemodialysis patients from red cell hemolysis.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 63 (the full 469-page PDF) before quoting it.