Middle Molecule Find

Rulebook › V-tags › V201

V201 RO: chemical analysis at installation and yearly

Written V0201 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 71 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

6.2.7 Reverse osmosis: (Chemical analysis: frequency)

Chemical analysis for the contaminants listed in 4.1.1 (Table 1) should be done when the RO system is installed, when membranes are replaced, and at not less than annual intervals thereafter to ensure that the limits specified in 4.1.1 are met (see Table 1). Chemical analyses should be done when seasonal variations in source water suggest worsening quality or when rejection rates fall below 90%.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

If your State has more stringent requirements, those must be followed to avoid potential noncompliance with 42 C.F.R. § 494.20..Accordingly, facilities who fail to comply with state and local laws would be cited at V100 or V101.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 71 (the full 469-page PDF) before quoting it.

Connected tags

Tags whose guidance points here