V594 Home HD testing per manufacturer and FDA labeling

Written V0594 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.100(c)(1)(v)(A) and (c)(1)(v)(B) · Appendix H Rev. 246, page 256 · checked 12 Nov 2026 · guidance updated Aug 17, 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§ 494.100(c) Standard: Support services. (1) …Services include, but are not limited to, the following: (v) …testing of the water and dialysate system in accordance with—
(A) The recommendations specified in the manufacturers’ instructions; and
(B) The system’s FDA-approved labeling for preconfigured systems designed, tested, and validated to meet AAMI quality (which includes standards for chemical and chlorine/chloramine testing) water and dialysate.

Read 42 CFR 494.100 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.100 (c)(1)(v)(A) and (B):

The home dialysis facility is responsible for monitoring the quality of the water/dialysate used by home hemodialysis patients, as required by the manufacturer’s recommendations and AAMI standards for the hemodialysis system. Water treatment systems for home hemodialysis patients must produce water that meets the AAMI standards and the requirements specified in § 494.40(a) of these regulations.

A chemical analysis of the product water must be done at the start of home treatment and at least once a year near the end of the usability of any disposable component, or when any modifications are made to the treatment components (other than the replacement of disposable components), to ensure that AAMI-defined maximum allowable chemical contaminant levels are not exceeded. If chemical analysis is not conducted as described here, refer to V201 for RO systems or to V206 for DI systems. According to AAMI, more frequent analysis than annual may be needed if there are seasonal variations in source water quality or if the source water is supplied from a well, as detailed in V593. When any repairs are made to water treatment equipment, the impact on water quality should be evaluated, and a chemical analysis should be performed if indicated.

Note: the requirements for in-center use of preconfigured systems are detailed at V276.