V205 Deionization: use as polisher or backup

Written V0205 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 75 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.2.8 Deionization (utilization as polisher or backup)

The usual application for a deionizer is as a polisher following reverse osmosis or as a standby process if the reverse osmosis system fails. Use of deionization as the primary means of purification in an outpatient facility is not recommended because of the inability of deionization and ultrafiltration to remove certain low-molecular-weight toxic bacterial products, such as microcystins.

Under no circumstances shall DI be used when the product water of the final bed has a resistivity below 1 megohm-cm.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.2.8 Deionization

The most common configuration for DI is to have two mixed beds in series, with resistivity monitors being placed downstream of each bed. Upon exhaustion of the first bed, reliance for water of sufficiently high resistivity shifts to the second bed, and dialysis operations may be continued for a short time (up to 72 hours) until a replacement bed is installed.