V206 Deionization: chemical analysis at installation and yearly
Written V0206 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
6.2.8 Deionization (chemical analysis: frequency)
When deionization is employed as the primary method for removing inorganic contaminants (reverse osmosis is not employed), or when deionization is necessary to polish RO-treated water, chemical analyses to ensure that the requirements of AAMI 4.1.1 (Table 1) are met should be performed when the system is installed and at annual intervals thereafter.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
Samples for chemical analysis should be drawn after the last treatment component; if the water treatment system includes DI as primary or polish, the routine chemical analysis samples should be taken after the DI. Taking the sample from the last patient treatment station would also meet this requirement.
Note: requirements for preconfigured systems, which may include DI, are given at V276.
Connected tags
Tags whose guidance points here
https://middlemolecule.com/rulebook/tags/v206/