V212 Distribution piping materials

Written V0212 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 82 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.3.3 Water distribution systems (no added burden)

Product water distribution systems shall be constructed of materials that do not contribute chemicals, such as aluminum, copper, lead, and zinc, or bacterial contaminants to the purified water.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.3.3 Water distribution systems

The choice of materials used for a water distribution system will also depend on the proposed method of disinfection. Whatever material is used, care should be taken to select a product with properties that provide the least favorable environment for bacterial proliferation, such as smooth internal surface. [See table below.]

Additional Guidance:

In the event a facility is using piping in the water distribution system which is not indicated as compatible with the disinfectant in use per Table 2, responsible staff (medical director, chief technician) should provide evidence that the disinfectant in use has been verified by the manufacturer of the disinfectant or the 510(k) licensed disinfection device as compatible with their distribution piping.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 82 (the full 469-page PDF) before quoting it.