V217 Hot water disinfection systems
Written V0217 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.3.4.3 Hot water disinfection systems
Refer to RD62:2001, 4.3.14 Hot water disinfection systems: When used to control bacterial proliferation in water treatment, storage, and distribution systems, the water heater of a hot water disinfection system should be capable of delivering hot water at the temperature and for the exposure time specified by the manufacturer.
5.3.4.3 Hot water disinfection systems
Hot water disinfection systems can be used only in systems constructed from heat-resistant materials, such as cross-linked polyethylene, polypropylene, and stainless steel (see [AAMI] 5.3.3).
The manufacturer’s instructions for using hot water disinfection systems should be followed. If no manufacturer’s instructions are available, the effectiveness of the system can be demonstrated by verifying that the system maintains a specified temperature for a specified time and by performing ongoing surveillance with bacterial cultures and endotoxin testing.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.3.4.3 Hot water disinfection systems
Hot water (≥80°C) may be used to control bacterial proliferation in water storage and distribution systems. Bacterial kill studies are not required.
Additional Guidance:
See Table 2 at V212 for acceptable heat-resistant materials.