V218 Hot water disinfection: monitoring and records

Written V0218 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 89 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

6.3.4 Bacterial control devices

6.3.4.3 Hot water disinfection systems: monitoring

Hot water disinfection systems should be monitored for temperature and time of exposure to hot water as specified by the manufacturer. Also, hot water disinfection should be performed at least as often as recommended by the manufacturer. The temperature of the water should be recorded at a point farthest from the water heater—that is, where the lowest water temperature is likely to occur…and measured each time a disinfection cycle is performed. A record that verifies successful completion of the heat disinfection should be maintained. Successful completion is defined as meeting temperature and time requirements specified by the equipment manufacturer.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

Records of use of hot water disinfection systems should include logs or recorded evidence to verify that the specified water temperature was maintained for the specified period of time in order to accomplish the intended disinfection.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 89 (the full 469-page PDF) before quoting it.