V219 Distribution system disinfection at least monthly
Written V0219 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
7 Strategies for bacterial control
7.1 General (disinfection and frequency)
Routine low-level disinfection of the pipes should be performed to control bacterial contamination of the distribution system. The frequency of disinfection will vary with the design of the system and the extent to which biofilm has already formed in existing systems, but disinfection must be performed at least monthly.
A mechanism should be incorporated in the distribution system to ensure that disinfectant does not drain from pipes during the disinfection period.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
7 Strategies for bacterial control
7.1 General
The strategy for controlling the proliferation of microorganisms in hemodialysis systems primarily involves proper system design and operation, and regular disinfection of water treatment system and hemodialysis machines. A key concept in ensuring compliance with water bacteriology standards is that disinfection schedules should be designed to prevent bacterial proliferation, rather than being designed to eliminate bacteria once they have proliferated to an unacceptable level (i.e., above the action level). With this strategy, monitoring levels of bacteria and endotoxins serves to demonstrate that the disinfection program is effective, not to indicate when disinfection should be performed. Gram-negative water bacteria, their associated lipopolysaccharides (bacterial endotoxins), and nontuberculous mycobacteria (NTM) most frequently come from the community water supply, and levels of those bacteria can be amplified depending on the water treatment system, dialysate distribution system, type of dialysis machine, and method of disinfection.
Two components of hemodialysis water distribution systems—pipes and storage tanks—can serve as reservoirs of microbial contamination. Hemodialysis systems frequently use pipes that are of larger diameter and longer than are needed to handle the required flow. Oversized piping slows the fluid velocity and increases both the total fluid volume and the wetted surface area of the system. Gram-negative bacteria in fluids remaining in pipes overnight multiply rapidly and colonize the wet surfaces, thus producing bacterial populations and endotoxin quantities in proportion to the volume and surface area. Such colonization results in the formation of protective biofilm that is difficult to remove once formed and that provides a barrier between the bacteria and germicide during disinfection.
Biofilms are communities of microorganisms attached to surfaces. They form just about anywhere a nonsterile fluid flows over the surface. Biofilm increases the ability of microorganisms to compete for nutrients and other resources. The complexity of biofilm depends on the degree of water or fluid movement and the availability of nutrients. Thicker biofilm, and usually a greater diversity of microorganisms, will form in slower-moving waters; in faster-moving waters, it is harder for microorganisms to become (and remain) attached to the surface, so biofilm formation takes longer. Organisms living within biofilm are shielded by an extracellular polymer or glycocalyx. This glycocalyx provides the bacteria with some protection from the action of disinfectants. Biofilm may exist throughout a hemodialysis distribution system. Once established in a distribution system or dialysis machine, biofilm can be difficult to eradicate. Bleach and ozone are generally the most effective agents for biofilm removal, and their use may be more efficacious if the pipes are treated first with a descaling agent. However, in some cases, complete or partial replacement of the distribution system may be the only way to eliminate biofilm.
Additional Guidance:
In order to minimize the risk of development of biofilm, every dialysis facility must disinfect their water distribution system at least monthly.
Delineation of what constitutes sufficient contact time of an EPA-registered disinfectant is typically determined by the manufacturer. All surfaces in the water distribution system must have sufficient contact, as instructed by the manufacturer, time with the disinfectant prior to its being rinsed from the system.