Middle Molecule Find

Rulebook › V-tags › V220

V220 Disinfecting the machine supply line

Written V0220 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 92 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

7 Strategies for bacterial control

7.1 General (disinfection of machine supply line)

Users should establish a procedure for regular disinfection of [the line between the outlet from the water distribution system and the back of the dialysis machine].

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

7 Strategies for bacterial control

7.1 General

For most dialysis machines, routine disinfection with hot water or with a chemical germicide connected to a disinfection port on the machine does not disinfect the line between the outlet from the water distribution system and the back of the dialysis machine. One approach is to rinse the dialysis machines with water containing germicide or hot water when the water distribution loop is disinfected. If this procedure is used with a chemical germicide, each dialysis machine should be rinsed and tested for the absence of residual germicide following disinfection.

Additional Guidance:

The line between the outlet from the water distribution system and the back of the dialysis machine is commonly referred to as the machine supply line. It is a hose that connects the dialysis machine to the treated water outlet. This hose should be disinfected at the same frequency as the water distribution loop is disinfected, i.e., monthly.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 92 (the full 469-page PDF) before quoting it.