V229 Concentrate batch records and testing before use
Written V0229 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.4.4.1 Mixing systems (permanent records)
6.4.1 Mixing systems (verification testing)
5.4.4.1 Mixing systems
In addition to container labeling, there should be permanent records of batches produced. These records should include the concentrate formula produced, the volume of the batch, the lot numbers of powdered concentrate packages, the manufacturer of the powdered concentrate, the date and time of mixing, any test results, the person performing the mixing, the person verifying mixing and test results, and the expiration date (if applicable).
6.4.1 Mixing systems
Acid and bicarbonate concentrates may be tested by using conductivity or by using a hydrometer.
Concentrates should not be used or transferred to holding tanks or distribution systems until all tests are completed. The test results and verification that they meet all applicable criteria should be recorded and signed by the individuals performing the tests.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
6.4.1 Mixing systems
Although not required, some manufacturers may provide allowable ranges for either the conductivity or the specific gravity of concentrates prepared from their powder. The use of pH as an indicator of proper dissolution is inappropriate for both acid and bicarbonate concentrates, because large variations in concentration do not produce significant changes in pH.
Additional Guidance:
The mixing logs should demonstrate complete documentation of this required information. Test results may include conductivity or specific gravity. Facility policy must stipulate the expected ranges for the test(s) used to verify correct mixing. Standards used to calibrate the instrument used to measure conductivity should encompass the expected results; for example, it would be unacceptable to use a “14” standard for an expected result of “70.” “Verifying mixing and test results” means the staff member performing this task checks the results against the expected ranges for the test and does not release mixtures for use that test outside those ranges.
Connected tags
Tags whose guidance points here
https://middlemolecule.com/rulebook/tags/v229/