V249 Proportioning ratio: machine labeling and monitoring
Written V0249 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.6 Dialysate proportioning (dialysis machine recalibration)
6.6 Dialysate proportioning
5.6 Dialysate proportioning (dialysis machine recalibration) [C]hanging from one proportioning ratio to another requires recalibration for some models of dialysis machines. Thus, for those machines, the type of concentrate should be labeled on the machine or clearly indicated by the machine display. It is strongly recommended that facilities configure every machine to use only one type of concentrate.
6.6 Dialysate proportioning
Dialysate proportioning should be monitored following the procedures specified by the equipment manufacturer. The user should maintain a record of critical parameters such as conductivity and approximate pH. When the user has specific requirements for monitoring dialysate proportioning, such as when dialysis machine settings are changed to allow the use of concentrates with a different proportioning ratio, the user should develop procedures for routine monitoring of dialysate electrolyte values.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.6 Dialysate proportioning
Some models of dialysis machines can use concentrates of only one type of proportioning ratio, but others may be set up or calibrated for use with concentrates of more than one proportioning type.
Additional Guidance:
The medical director and responsible staff should be knowledgeable of the mixing ratio the machines are set up to use and all dialysate supplies in the facility must match that ratio.
Rarely, a facility may have machines set for different ratios; this is a risky practice and would require very close monitoring to prevent a mismatch of supplies and machines. If machines are available for different ratios in the same facility, each machine should be clearly labeled for the applicable ratio and supplies for the different ratios should be segregated and labeled clearly to avoid mismatch. The medical director should be aware of this practice and be involved in quality control to avoid any patient consequence from potential mismatch of supplies and machines.
If machines are changed from one ratio to another, responsible staff members should be able to describe how they verified the machine functioned correctly after the change was made and how they monitored the dialysate electrolyte values.