V250 Conductivity and pH before each treatment
Written V0250 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.6 Dialysate proportioning (dialysate conductivity and pH measurement) [I]t is necessary for the operator to follow the manufacturer’s instructions regarding dialysate conductivity and to measure approximate pH with an independent method before starting the treatment of the next patient.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.6 Dialysate proportioning
Injuries related to improper dialysate are rare, but they can and do happen when all procedures are not followed. Frequently, when the error occurs, several patients have been exposed before the facility recognizes the mistake. For example, because one of the concentrates is quite acidic and the other is basic, connecting the wrong concentrates to the machine could result in dialysate that could harm the patient.
Even though a single concentrate is used to prepare acetate dialysate, conductivity and pH should be checked, because certain mix-ups involving acid concentrate and other chemicals can result in an acceptable conductivity and very low pH.
Additional Guidance:
Each machine must be tested for pH using a handheld meter or other appropriate testing device (i.e., adequately sensitive testing strips) before every dialysis treatment and whenever a different composition of acid concentrate is used. If the dialysis machine manufacturer requires testing of conductivity, this must also be tested using an independent testing device prior to each treatment and before using a different composition of acid concentrate during the same treatment. The facility must have set limits for the allowable variability of the handheld values from the machine readings. As new technology evolves, the manufacturer’s guidance related to this requirement should be followed.