V252 Water cultures: frequency and sample collection
Written V0252 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
7.2 Microbial monitoring methods
7.2.1 General
7.2.2 Sample collection
7.2.1 General
Culture water …weekly for new systems until a pattern has been established. For established systems, culture monthly unless a greater frequency is dictated by historical data at a given institution.
Monitoring can be accomplished by direct plate counts, in conjunction with the measurement of bacterial endotoxin.
7.2.2 Sample collection
Water samples should be collected directly from outlet taps situated in different parts of the water distribution system. In general, the sample taps should be opened and the water should be allowed to run for at least 60 seconds before a sample is collected in a sterile, endotoxin-free container. A minimum of 50 mL of water, or the volume specified by the laboratory performing the test, should be collected. Sample taps should not be disinfected.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
7.2.2 Sample collection
All new sterile plastic ware is endotoxin-free because of the high temperatures involved in the manufacturing process.
Additional Guidance:
Bacterial colony counts may be done by an outside lab (thus the reference to direct plate counts) or by the use of dip samplers, as described at V256. If a pattern of results greater than the accepted limits is identified, the frequency of cultures (and subsequent disinfection) should be increased to reach a frequency that ensures the system is being maintained below the allowed contamination level between scheduled sampling. Refer to V255.
Connected tags
Tags whose guidance points here