Middle Molecule Find

Rulebook › V-tags › V270

V270 Chlorine breakthrough: fix and confirm, or stop dialysis

Written V0270 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.40(b)(2)(ii)(A) · Appendix H Rev. 246, page 124 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§ 494.40(b)(2)(ii) - If the test results from the last component or carbon tank are greater than the parameters for chlorine or chloramine specified in paragraph (b)(2)(i) of this section the facility must—
(A) Immediately take corrective action to bring chlorine or chloramine levels into compliance with paragraph (b)(2)(i) of this section and confirm through testing that the corrective action has been effective, or terminate dialysis treatment to protect patients from exposure to chlorine/chloramine; ]\

Read 42 CFR 494.40 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.40(b)(2)(ii)(A)

“Corrective action” here could include backwashing carbon tanks, rebedding or replacement of the tanks, or addition of an adjunct system such as chemical injection to address an extremely high chlorine or chloramine load from the municipal supplier. In any case, testing must confirm acceptable levels of chlorine/chloramine before dialysis treatments can resume. It could be necessary to reschedule patient treatments, or for patients to be transferred to another facility for one or more treatments.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 124 (the full 469-page PDF) before quoting it.

Connected tags

Tags whose guidance points here