V314 Water for reprocessing: quality and testing
Written V0314 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)
7.1 Water systems
The system providing water for reprocessing shall meet all of the requirements for pressure and flow rate for operating the reprocessing equipment under minimal and peak load conditions. Product water used for rinsing, cleaning, filling, and diluting the germicide shall be shown to comply with the chemical and microbiological quality requirements [specified in these regulations].
Water bacteriology monitoring shall be carried out where the dialyzer is connected to the reuse system or as close as possible to that point.
11.4 Germicide
11.4.1.5 Water quality monitoring
The water used to rinse and clean dialyzers and dilute the germicide should be tested for bacterial contamination and pyrogens according to the requirements [of these regulations] before a reprocessing program is undertaken. Once dialysis with the reprocessed hemodialyzers has begun, testing for bacterial contamination should be frequent (e.g., weekly). Less frequent testing, but not less than monthly, may be appropriate if there is a documented history of at least 3 months of results consistently below the required levels.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
The product water chemical and microbiological requirements outlined in this section are the same as those in ANSI/AAMI RD52:2004, incorporated by reference in these regulations under the Condition for water and dialysate quality.
Water samples for microbial and endotoxin testing must be routinely taken each month from the water supplying each reprocessing system, as close as possible to the point where the dialyzer would be connected to the system. If more than one automated reprocessing system is in use, the water supply to each system must be monitored monthly.
New facilities or facilities which add dialyzer reprocessing should validate the safety of the water supply to the reprocessing system by testing for bacteria (microbial content) and pyrogens (endotoxins) weekly for at least 3 months, and at least monthly thereafter. The sites to be tested include the water supply to the sinks used for rinsing dialyzers, to outlets used for mixing germicide, and to each reprocessing system.
Note: If water quality samples do not consistently meet AAMI standards, i.e., at least 3 consecutive months of testing results that meet required levels, the water quality testing is expected to be performed more frequent than monthly.
https://middlemolecule.com/rulebook/tags/v314/