V334 Header cleaning and disinfection
Written V0334 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)
11.4.1.2 Dialyzer header cleaning and disinfection
The cleaning and disinfection of the header space should be done only when necessary and only before the dialyzer is reprocessed. The manufacturer’s instructions should be followed. Header caps and O-rings shall be kept with their respective dialyzers.
If the header cap is removed to clean the header space, cleaning shall be done with water meeting the requirements of these regulations related to allowable bacterial and endotoxin levels.
Once the O-ring and the header cap are cleaned and before they are reassembled at the end of the dialyzer, they should be disinfected. The disinfectant shall not be rinsed and shall be allowed to remain on the dialyzer components as they are reassembled. If any cracking of the header occurs, the process should be evaluated.
If the header space is cleaned with the header cap in place, it is necessary to ensure that the end of the fiber bundle is not damaged. If water is used, it shall meet the requirements of these regulations.
If automated equipment is used, the manufacturer’s instruction for use shall be followed.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
ANSI/AAMI RD47:2002
11.4.1.2 Dialyzer header cleaning and disinfection
Over tightening the header caps may cause damage to the cap, and under tightening the cap may cause blood leaks.
AAMI Rationale for the Development and Provisions of this Recommended Practice
A.11.4.1.2 Dialyzer header cleaning and disinfection
The practice of header removal to remove clotted material has increased over the years. Removing the header allows the user to remove the clotted material from the end of the fiber bundle and the O-ring header assembly. The method of removing the clotted material has been a concern. Some facilities use running water (AAMI quality) to remove the clotted material, whereas others use 4x4s or instruments to scrape away the clotted material. The main concerns of using 4x4s or instruments to scrape away the clotted material are (1) infection, (2) plugging of fibers, and (3) damage to the end of the fiber bundle.
In the past, removing the headers was associated with reported incidents of bacterial and pyrogenic reactions in patients. The patient reactions ceased to occur when the headers were disinfected by dipping the O-ring, header, and end of the dialyzer into the appropriate disinfectant. The research on this problem pointed to a double-fault failure system: 1) the bacteria seemed to be coming from a contaminated water source, and 2) the bacteria were not killed by the normal disinfection process. Dipping the dialyzer corrected that situation.
Several concerns are raised when the headers are not removed and the user attempts to clear the header space of clots. These concerns include infection and damage to the end of the fiber bundle. A multitude of items are used to clean the header space, including water sprays, paper clips, tie wraps, and the like. With water sprays, the possibility of contaminated water always exists. Other items that are inserted can damage the end of the fiber bundle. If the item inserted into the dialyzer is not disinfected between uses, it can cause bacterial transmission; however, the dialyzer is usually disinfected after the header space is cleaned. Automated header cleaning devices are commercially available.
Additional Guidance:
According to the Centers for Disease Control and Prevention (CDC), if the ends of the dialyzer (header caps) are removed for cleaning, only a stream of AAMI quality water may be used to clean blood clots, etc. from the exposed ends and the header caps of dialyzers.
If the header caps are removed during reprocessing, facility staff must ensure that the caps, O-rings, and the ends of the dialyzer are immersed or saturated with germicide prior to reassembly, and that the components are reassembled wet with germicide. Cleaning of header caps and dialyzer ends must be closely audited for any breaks in technique which could put the patient at risk.