Middle Molecule Find

Rulebook › V-tags › V333

V333 Pre-cleaning with water that meets the limits

Written V0333 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 153 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

11.2 Rinsing/cleaning

11.2.3

Pre-cleaning the dialyzer (rinsing and cleaning) shall be done with a fluid or fluids made with water that meets the requirements of these regulations related to allowable bacterial and endotoxin levels.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

All water used for rinsing and reprocessing the dialyzer interior must meet the requirements of AAMI RD52:2004 regarding allowable bacterial and endotoxin levels. Refer to § 494.140(a) (V178) for details on microbial counts and endotoxin concentration levels . The interior of the dialyzer should never be exposed to tap water. The facility must monitor the water supply to the reprocessing station(s) for bacteria and endotoxins.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 153 (the full 469-page PDF) before quoting it.