Rulebook › V-tags › V347

V347 Inspection before use

Written V0347 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 167 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)

12.1 Visual inspection (dialyzer inspection prior to use)

The dialyzer should be inspected before it is prepared for use. Completion of this inspection should be recorded in the reprocessing record (see [AAMI] 4.2), along with the signature or other unique means of identifying the person completing the inspection. The inspection should include the following:

  • The reprocessed dialyzer shall be legibly labeled with the information recommended in [AAMI] 10.3.
  • There should be no indication of structural damage or tampering with the dialyzer.
  • The ports of the dialyzer should be properly capped.
  • The presence of germicide in the dialysate and blood compartments, including headers, should be confirmed, and there should be no evidence of leakage from the ports or other portions of the dialyzer.
  • The duration and conditions of storage should be appropriate for the agent or method used to sterilize or disinfect the dialyzer; and
  • The cosmetic appearance of the dialyzer should be aesthetically acceptable to the staff and the patient.

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.11.5 Inspection

The [AAMI RDD] Committee recognized that the patient should be included in the aesthetic evaluation of the hemodialyzer.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 167 (the full 469-page PDF) before quoting it.