V348 Matching the dialyzer to the patient
Written V0348 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR 494.50(b)(1)
12.2 Verification of patient identification
Except in the case of home dialysis, two people should check that the first and last names on the dialyzer [and any other appropriate identifying information] correspond to the identifying information on the patient’s permanent record. If possible, one of the people checking identification should be the patient. Completion of this step shall be recorded, along with the signature or other unique means of identifying the person verifying patient identification.
NOTE—This step may be done later in the procedure but shall precede initiation of dialysis.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(1)
Standard of practice requires the final check be done when the patient is present for that treatment. If possible, patients should be encouraged to check their dialyzers for their names. Patients who sign the treatment record for this item should understand what their signature means. If patients are unable to identify their dialyzers, two staff should do so and record the verification prior to initiating the dialysis. Note that some facility policies require two staff to verify patient/dialyzer identification for all reused dialyzers.