The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
4.1.2 Bacteriology of water: Product water used to prepare dialysate or concentrates from powder at a dialysis facility, or to process dialyzers for reuse, shall contain a total viable microbial count lower than 200 CFU/mL and an endotoxin concentration lower than 2 EU/mL
The action levels for the total viable microbial count in the product water shall be 50 CFU/mL, and for the endotoxin concentration, 1 EU/mL. If those action levels are observed in the product water, corrective measures shall promptly be taken to reduce the levels.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
AAMI Rationale for the Development and Provision of This Recommended Practice
A.4.1.2 Bacteriology of water
When ANSI/AAMI RD5:1981 was initially developed, it was generally considered that the water used to prepare dialysate did not require sterilization. Studies have demonstrated that the incidence of pyrogenic reactions is directly related to the number of bacteria in dialysate. It is also known that a dialysate delivery system can amplify the level of bacteria in the water used to supply the system. Those studies provided the rationale for setting a recommended maximum concentration of 200 bacteria per mL in the water used to prepare dialysate.
Several groups of investigators have convincingly demonstrated that pyrogenic reactions are caused by lipopolysaccharides, also known as endotoxins, associated with Gram-negative bacteria. Gram-negative water bacteria can multiply rapidly in the chemically pure water used to supply hemodialysis systems. It has also been demonstrated clearly that endotoxins and endotoxin fragments can cross both low-flux and high-flux hemodialysis membranes.
Because 48 hours can elapse between sampling water to determine microbial contamination and receiving results, and because bacterial proliferation can be rapid, action levels for microbial counts and endotoxin concentrations are included in these regulations. Those action levels allow the user to initiate corrective action before levels exceed the recommended maximum levels. Unlike cultures, endotoxin testing does not require extended incubation times. Endotoxin testing, when performed in the dialysis facility, can yield results in approximately 1 hour, thereby eliminating the long delay between sampling and obtaining a result.
During the development of this recommended practice, the AAMI RDD Committee was tasked with recommending levels of bacteria and endotoxins above which the water should not be used for dialysis applications. In making the recommendations outlined in ANSI/AAMI RD52:2004, Section 4.1.2, the AAMI RDD Committee understood that dialysis would continue at contaminant levels above the action level but below the recommended maximum level. Establishing a recommended maximum level of contamination at which dialysis should be stopped immediately is difficult because the risk of adverse events, such as pyrogenic reactions, must be balanced against the risks of uremia if a patient is not dialyzed. The balance between these two risks will depend on the level of contamination, the time of exposure, and the patient's medical condition. Because this balance will almost certainly vary from circumstance to circumstance, the AAMI RDD Committee felt that there was insufficient data on which to base levels of bacteria and endotoxins above which dialysis should not be performed. The final decision of whether to discontinue dialysis rests with the medical director of a facility. Whatever decision is made, the AAMI RDD Committee recommends that the water treatment and distribution system be disinfected promptly any time the levels of bacteria or endotoxins exceed the action levels recommended in AAMI 4.1.2. Additionally, it may be prudent to discontinue dialyzer reuse if the levels of bacteria or endotoxins exceed the recommended maximum levels outlined in AAMI 4.1.2, as the water is introduced directly into the blood compartment of the dialyzer.
Additional Guidance:
If the facility reaches the “action” level, remedial action is required. Action could be to repeat a culture, particularly if only one in a set of cultures was above the action limit. Action could also include disinfecting the system and repeating cultures at several sites.
Some states require this testing to be done in a laboratory approved for the analysis of potable water.